Gas Analysers
Digital gas flow analyser for connection to fermenters or as stand-alone laboratory instruments: LAMBDA OXYMETER, CARBOMETER & METHAMETER
LAMBDA OXYMETER
The gas flow analyser LAMBDA OXYMETER is a flow-through optical oxygen meter for applications where the Oxygen concentration in the gas flow needs to be determined. The LAMBDA OXYMETER is designed for the Minifor2Bio fermentation system, however, the laboratory measurement instrument can also be used separately as a stand-alone instrument.
LAMBDA OXYMETER uses a factory-calibrated Oxygen sensor which measures ambient oxygen levels using the principle of fluorescence quenching by oxygen, also known as luminescence quenching.
The luminescence quenching principle measures the partial pressure of oxygen (ppO2. The internal barometric pressure sensor measurement along with ppO2 measurement allows the oxygen concentration (O2%) value to be calculated.
The in-built sensor benefits from a low-power operation, traditionally associated with electrochemical and galvanic sensors, while providing a much longer lifetime due to the non-depleting sensing principle.
Features of LAMBDA OXMETER gas flow analyser
- Luminescence–based optical technology
- Temperature compensation
- Barometric pressure compensation
- 100% RoHS compliant sensor
- Solid state sensor
- Factory calibrated
- Long life due to non-depleting sensing principle
- 5” TFT IPS 320 x 240 pixelswith capacitive touch panel
- USB 1.1/2.0 communication interface
- Based on 32 bit 120 MHz microprocessor
- Optionally powered through the REMOTE connector from the Minifor2Bio unit
- Meets safety standards and conformity for laboratory use
OXYMETER | ||
---|---|---|
Type | Oxygen flow-through meter | |
Oxygen range | 0–25% O2 | |
Oxygen pressure | 0–300 mbar ppO2 | |
Responce time | < 30 s (typically) | |
Accurancy | ppO2 | < 2% FS |
Temperature | Indication only | |
Pressure | ± 5 mbar | |
O2 | Determinated by ppO2 and pressure accuracy | |
Resolution | ppO2 | 0.1 mbar |
Temperature | 0.1 °C | |
Pressure | 1 mbar | |
O2 | 0.01 % | |
Sensor lifetime | > 5 years | |
Operating pressure | 0.5–2 bar | |
Maximum flow rate | 1.0 l/min | |
Barometric pressure | 500–1200 mbar | |
Interface | USB 1.1/2.0 & REMOTE (CAN BUS) | |
Display | 3.5” TFT IPS with 320 x 240 pixels resolution Viewing angles ± 70° |
|
Conformity | DIRECTIVE 2014/35/EU DIRECTIVE 2014/30/EU |
|
Technical Standards | EN 61000-6-1, EN-61000-6-4, EN 61010-1:2010/A1:2019/AC:2019-04 EN 61326-1:2013 |
|
Dimensions (W x H x D) | 104.4 mm x 94.7 mm x 105 (114.5) mm | |
Weight | < 1 kg | |
Operating temperature | 0–40 ⁰C | |
Operating humidity | 0–90%, no icing or condensation | |
Power supply | Plug-in power adapter 90–240V AC 50/60 Hz 12V / 1.0A 12W barrel jack 5.5/2.1 |
LAMBDA CARBOMETER – gas flow ANALYSER for CO2
The LAMBDA CARBOMETER is a flow-through fast response CO2 analyser for applications where the Carbon Dioxide concentration in the gas flow need to be determined. The LAMBDA CARBOMETER is designed for the Minifor2Bio fermentation system, however, it can also be used separately as a stand-alone instrument.
The in-built Carbon Dioxide sensor integrates unique mid infrared light source and detector technology with innovative optics and signal processing.
The sensor offers leading-edge technology, ultra-fast response time, low power requirements, and a wide measurement range of up to 100% CO2.
Features of LAMBDA CARBOMETER gas flow analyser
- Fast response
- Mid infra-red light source technology
- Measures up to 100% CO2 concentration
- Solid-state NDIR optical technology
- Vibration and shock-resistant sensor
- Non-heating technology
- Ability to operate in harsh environmental conditions
- Sensor lifetime > 15 years
- 5” TFT IPS 320 x 240 pixels with capacitive touch panel
- USB 1.1/2.0 communication interface
- Based on a 32 bit 120 MHz microprocessor
- Optionally powered by the REMOTE connector from the Minifor2Bio unit
- Meets safety standards and conformity for laboratory use
CARBOMETER | |
---|---|
Type | Carbon Dioxide flow-through meter |
Measurement ranges | 0–20% or 0–100% |
Responce time | < 10 s (typically) |
Accurancy | 0–20% ± (70 ppm + 5%) 0–100% ± (300 ppm + 5%) |
Sensor Sample Method | Solid-state LED NDIR Diffusion |
Responce time | < 30 s (diffusion limited) |
Pressure Dependence | 500 mbar–2 bar |
Interface | USB 1.1/2.0 & REMOTE (CAN BUS) |
Display | 3.5” TFT IPS with 320 x 240 pixels resolution Viewing angles ± 70° |
Conformity | DIRECTIVE 2014/35/EU DIRECTIVE 2014/30/EU |
Technical Standards | EN 61010-1:2010/A1:2019/AC:2019-04 EN 61326-1:2013 |
Dimensions (W x H x D) | 104.4 mm x 94.7 mm x 105 (114.5) mm |
Weight | < 1 kg |
Operating temperature | 0–40 ⁰C |
Operating humidity | 0–90%, no icing or condensation |
Power supply | Plug-in power adapter 90–240V AC 50/60 Hz 12V / 1.0A 12W barrel jack 5.5/2.1 |
METHAMETER
One of the gas flow meters used in anaerobe fermentation is the METHAMETER. The LAMBDA METHAMETER flow-through linearized and temperature compensated CH4 analyser for applications where the methane concentration in the gas flow need to be determined. The LAMBDA METHAMETER is designed for the Minifor2Bio fermentation system, however, it can also be used as a stand-alone instrument.
The in-built volume methane (CH4) sensor with auto range option, additionally plus 0–2% Propane, provides a measurement of Methane over the full range of 0 to 100% volume with a resolution of 0.01 % for 0–10% methane concentration range and a resolution of 0.1% for 10–100% methane concentration range.
Features of LAMBDA METHAMETER gas flow analyser
- Measurement ranges with auto-ranging option
- Fully temperature compensated
- Zero function
- Sensor ATEXT and UL standards certification
- Infrared technology
- 5” TFT IPS 320 x 240 pixels with capacitive touch panel
- USB 1.1/2.0 communication interface
- Based on a 32 bit 120MHz microprocessor
- Optionally powered by the REMOTE connector from the Minifor2Bio unit
- Meets safety standards and conformity for laboratory use
METHAMETER | |
---|---|
Type | LAMBDA METHAMETER flow-through meter |
Measurement ranges and resolutions | 0–10% (0.01% resolution) 0–100% (0.1% resolution) |
Accurancy | ± 2% |
Zero repeatability | ±0.05% |
Long term zero drift | ±0.05% |
Responce time | < 30 s |
Sensor Sample Method | IR sensor |
Responce time | < 30 s (diffusion limited) |
Pressure | Accuracy limits are maintained at pressures within ± 5% of the calibration pressure |
Interface | USB 1.1/2.0 & REMOTE (CAN BUS) |
Display | 3.5” TFT IPS with 320 x 240 pixels resolution Viewing angles ± 70° |
Conformity | DIRECTIVE 2014/35/EU DIRECTIVE 2014/30/EU |
Technical Standards | EN 61010-1:2010/A1:2019/AC:2019-04 EN 61326-1:2013 |
Dimensions (W x H x D) | 104.4 mm x 94.7 mm x 105 (114.5) mm |
Weight | < 1 kg |
Operating temperature | 0–40 ⁰C |
Operating humidity | 0–90%, no icing or condensation |
Power supply | Plug-in power adapter 90–240V AC 50/60 Hz 12V / 1.0A 12W barrel jack 5.5/2.1 |